Dry-well calibrators are portable instruments for the simple on-site calibration of temperature measuring instruments. A temperature dry-well calibrator consists of an electrically heated or cooled metal block, an accurate temperature control and an internal reference with indicator for determining the block temperature. In the metal block is a bore for accepting the insert. This exchangeable insert is supplied with one or more bores with individual diameters and serves for the adaption of the test items.
With on-site calibration, the thermometer to be tested can be taken directly out of its thermowell in the plant, without needing to interrupt the electrical connections to the evaluation system, and inserted in to the appropriate bore of the insert. Here it will generally be ensured that the bore diameter, bearing in mind the thermal expansion of the calibration item, is selected to be as tight as you possibly can, since any air gap between your test item and the insert impairs heat transfer. Therefore, the bore diameter must be no more than 0.5 mm higher than the external diameter of the test item.
Structure of a dry-well calibrator
The required temperature with which the test item ought to be calibrated is regulated with the calibrator. Frequently temperature probes are calibrated at the respective process temperature. Once a thermal equilibrium has been set between the thermometer being tested and the calibrator, the measuring deviation of the test item could be recorded. This corresponds to the temperature difference between its temperature and the reference temperature displayed on the calibrator. The full total measurement uncertainty comprises of the accuracy of the calibrator and the short-term stability of the test item. According to the temperature range and the properties of the calibration item, calibration is carried out with a measurement uncertainty of 0.1 K to 3 K.
What else ought to be taken into account?
The minimum immersion depth for dry well calibrators is 70 mm. That is driven by the axial gradient, since heated air rises from underneath up. If the minimum immersion depth not be reached, an external reference thermometer must be used for calibration. pressure gauge is then placed on the same level as the object to be calibrated and the outcomes are compared with one another.
Recommended guideline for the immersion depth of a test item
Diameter of the test item x 15 or for a known sensitive length: sensitive length + diameter of the test item x 10
Note
Further information on our portable temperature calibrators and services can be found on the WIKA website.
โซล่าเซลล์คุณภาพดี: เพื่อการใช้ประโยชน์จากพลังงานที่มีคุณภาพสูง
1. โซล่าเซลล์คุณภาพดี: เป็นอะไรและทำไมคุณต้องใช้? 1.1 โซล่าเซลล์คืออะไร? โซล่าเซลล์คือเทคโนโลยีที่ใช้แผงโซล่าเซลล์ในการแปลงพลังงานจากแสงอาทิตย์เป็นไฟฟ้าได้ โดยมีหน้าที่ส่วนหน้าของแผงโซล่าเซลล์ (solar panel) ที่ถูกออกแบบมาเพื่อรับแสงอาทิตย์และแปรส่วนที่ได้รับมาเป็นไฟฟ้า 1.2 การใช้โซล่าเซลล์มีประโยชน์อย่างไร? การใช้โซล่าเซลล์มีประโยชน์อย่างมากมาย ไม่เพียงแต่จะช่วยลดต้นทุนในการใช้ไฟฟ้าและอนุรักษ์พลังงาน แต่ยังช่วยลดการปล่อยก๊าซเรือนกระจก (greenhouse gas) ที่เป็นสาเหตุของการเปลี่ยนแปลงสภาพภูมิอากาศ นอกจากนี้โซล่าเซลล์ยังมีความคุ้มค่าในระยะยาว เพราะจะช่วยปรับปรุงคุณภาพชีวิตของเราให้ดียิ่งขึ้น